Oxygen Inhalator/ Oxygen Flowmeter Use for Patients

Model Number: Dsw 905b905b, 906b
Structure: Spring-Piston
Connection Thread: G5/8, Cga540
Wet The Cup: 200ml
Humidity Conditioning: 200ml
Applicable Gas: Oxygen
Customization:

Contact Supplier

Diamond Member Since 2006

Suppliers with verified business licenses

Trading Company

Basic Info.

Model NO.
DSW 905B905B, 906B
Input Pressure
15MPa
Output Pressure
0.2-0.3MPa
Safety Valve
0.35±0.05MPa
Material
Brass
Transport Package
Carton
Trademark
DSW
Origin
China
HS Code
9019200000
Production Capacity
50000PCS/Year

Product Description

Type Brand Name Model Number Place of Origin  Applicable Gas Input Pressure Output Pressure Safety valve The Range of Flow
Breathing Apparatus  DSW DSW 905B905B,906B DSW China Oxygen    15MPa 0.2-0.3MPa   0.35±0.05MPa  1-15L/min
Connection Thread  Structure Wet the cup Humidity Conditioning Packaging Details  Dimension G.W N.W Material
G5/8,CGA540  Spring-piston 200ML  200ML 10PCS/CTN  53*44*18cm  11.5Kg 10Kg   brass

Oxygen inhalator/ oxygen flowmeter use for patients  

Outside thread G5/8,
Flow 0-10L/min or 0-15L/min
Output pressure 0.2-0.3MPa,
Input pressure 150bar
 
gas inlet types: 
1.gas -specific  inlet  probe conforming to BS ,AMERICAN, GERMAN, FRENCH 
2.1/8'' NPT female-threadde inlet 
3.Bullnose nipple with 3/8'' BSPNUT.   Other country-specific direct and remote-fitting configurations are 
available- contact the office. 
gas outlet types: 
1.body has female thread -use BMI 'universal'tubing nipple 
2.body has a 9/16'' UNFmale thread -use tubing nipple(oxygen) 
mounting: 
1.direct-probe fits into wall-mounted gas -specific socket(antirotation device maintains flowmeter upright) 
2.remote-kit available with 2m colour-coded flexible hose,and a gas-specific connector the flowmeter is mounted 
remotely from  the gas source on: 
a)wall/machine-mouted V-plate 
b)medical rail. 
NOTE:the rail-mounting kit utilises a "universal' rail clamp,allowing attachment to BMI,ohmeda,vickers"modura",
din rail section. 
operation and use 
2.service & maintence 
do dot disassemble when connected to a source of pressure 
caution 
repairs must only be undertaken by suitably trained personnel. 
the pre-use checks in section 2, and the leakage check in section 3.2 must be carried out after any disassembly 
3.tube assembly 
outer tube 
if the cover tube is cracked, withdraw the unit from service and renew: 
1.disconnect the flowmeter from the pressure source 
2.open the flow control valve to relieve any contained pressure. 
3.hold the unit vertically (to retain the flow tube and ball )unscrew the cover tube and lift off. 
calibrated tube and indicator ball 
4.lift off the flow tube , taking care not to let the ball roll out 
5.remove the tube connector adaptor and the cover tube o-ring 
6.fit a new cover tube o-ring at the base of the body. 
7.fit a new tube adaptor. Check orientation,as illustrated. 
8.fit the flow tube , and ball over the adaptor. 
9.position the tube with the scale towards the front of the unit. 
10.fit the outer tube,and screw clockwise into the flowmeter body. 
4. environmental conditions 
store/operate between 5 and 25 degree 
protect exposed o-rings from light 
Oxygen Inhalator/ Oxygen Flowmeter Use for Patients
Oxygen Inhalator/ Oxygen Flowmeter Use for Patients
Oxygen Inhalator/ Oxygen Flowmeter Use for Patients

 

Send your message to this supplier

*From:
*To:
avatar Mrs. Cathy
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Contact Supplier

Diamond Member Since 2006

Suppliers with verified business licenses

Trading Company
Number of Employees
23
Year of Establishment
2004-01-09